Invacare Orion Service Manual page 43

Hide thumbs Also See for Orion:
Table of Contents

Advertisement

®
Invacare
- Orion
·
Remove screws (1) on potentiometer.
·
Replace potentiometer
·
Reconnect potentiometer plug (1) with
operating console.
Setting the potentiometer to the middle
position:
·
Loosen the bolts (1) on the drive lever
mounting.
·
Twist the potentiometer shaft (2) until the
same resistance (approx. 5kOhm) is
measured between the central connection
cable (3) and both outputs (4).
·
Retighten the bolts on the drive lever
mounting.
·
Check settings and repeat the adjustment if
necessary.
·
Reassembly of the operating console takes
place in reverse order
·
Check all functions (trial run).