4.2
Piping ST 3000 Transmitter,
Installing flange
adapter
ATTENTION
38
Table 14 gives the steps for an optional flange adapter on the process
head.
Slightly deforming the gasket supplied with the adapter before you
insert it into the adapter may aid in retaining the gasket in the groove
while you align the adapter to the process head. To deform the gasket,
submerse it in hot water for a few minutes then firmly press it into its
recessed mounting groove in the adapter.
Table 14
Installing Flange Adapter
Step
1
Insert filter screen (if supplied) into inlet cavity of process head.
Carefully seat Teflon (white) gasket into adapter groove.
2
Thread adapter onto 1/2-inch process pipe and align mounting holes
3
in adapter with holes in end of process head as required.
4
Secure adapter to process head by hand tightening 7/16-20 hex-
head bolts.
Example - Installing adapter on process head.
ATTENTION
bolts prior to threading them into the process head.
Evenly torque flange adapter bolts to a torque of 27,1 Nm +/- 1,4 Nm
5
(20 ft lbs +/- 1.0 ft lbs)
ST 3000 Release 300 Installation Guide
Continued
Action
Filter Screen
Teflon Gasket
Flange Adapter
7/16 x 20 Bolts
Apply an anti-seize compound on the stainless steel
Process
Head
21011