Optional AC Voltage Performance Verification Test
Configuration: AC Volts (CONFigure[:VOLTage]:AC)
1 Select the AC Voltage function.
2 Select each range in the order shown below. Provide the indicated input
Table 11
Vrms
[1] Select 5½- digit measurement resolution
34405A User's and Service Guide
voltage and frequency. Compare measurement results to the appropriate
test limits shown in the table. (Be certain to allow for appropriate
source settling.)
Optional AC Voltage Performance Verification Test
Input Frequency Function
1V
45Hz
1V
1kHz
1V
10kHz
1V
30kHz
1V
100kHz
10V
1kHz
1V
1kHz
0.1V
1kHz
[1]
Range
AC Voltage
1V
1V
1V
1V
1V
10V
10V
10V
Performance Tests and Calibration
Error from Nominal
1 year
±11mV
±3mV
±3mV
±11mV
±32mV
±30mV
±12mV
±10.2mV
4
79