Optional Ac Voltage Performance Verification Test - Agilent Technologies 34405A User's And Service Manual

5 1/2 digit multimeter
Hide thumbs Also See for 34405A:
Table of Contents

Advertisement

4
Performance Tests and Calibration

Optional AC Voltage Performance Verification Test

84
Configuration: AC Volts (CONFigure[:VOLTage]:AC)
1 Select the AC Voltage function.
2 Select each range in the order shown below. Provide the
indicated input voltage and frequency. Compare
measurement results to the appropriate test limits shown
in the table. (Be certain to allow for appropriate source
settling.)
Table 11
Optional AC Voltage Performance Verification Test
Vrms Input Frequency Function
1V
45Hz
1V
1kHz
1V
10kHz
1V
30kHz
1V
100kHz
10V
1kHz
1V
1kHz
0.1V
1kHz
[1] Select 5½- digit measurement resolution
[1]
Range
AC Voltage
1V
1V
1V
1V
1V
10V
10V
10V
34405A User's and Service Guide
Error from Nominal
1 year
±11mV
±3mV
±3mV
±11mV
±32mV
±30mV
±12mV
±10.2mV