Analog Bandwidth - Maximum Frequency Check

C A U T I O N
Ensure that the input voltage to the oscilloscope never exceeds ±5 V.
Let the oscilloscope warm up before testing
The oscilloscope under test must be warmed up (with the oscilloscope application running) for at least
30 minutes prior to the start of any performance test.
Specification
Analog Bandwidth (-3 dB)
DSO/DSA91304A
DSO/DSA91204A
DSO/DSA90804A
DSO/DSA90604A
DSO/DSA90404A
DSO/DSA90254A
Equipment Required
Description
Microwave CW
Generator
Power Splitter
Power Meter
Power Sensor
Microwave Cable
SMA Adapters
SMA to BNC Adapter
12.0 GHz, 11.8 GHz at
5 mV/div
12.0 GHz, 11.8 GHz at
5 mV/div
8.0 GHz
6.0 GHz
4.0 GHz
2.5 GHz
Critical Specifications
Maximum Frequency ≥ 14 GHz
Power range: -20 dBm to +16 dBm into 50Ω
Output resistance = 50Ω
2 Resistor Power Splitter
Max Frequency ≥18 GHz
Agilent E-series with power sensor compatibility
Maximum Frequency ≥ 14 GHz
Power range: -24 dBm to +16 dBm
50Ω Characteristic Impedance
3.5 mm (m) to 3.5 mm (m) SMA connectors
Max Frequency ≥18 GHz
3.5 mm (m) to 3.5 mm (m) SMA
3.5 mm (f) SMA to Precision BNC (No Substitute)
Chapter 3: Testing Performance

Analog Bandwidth - Maximum Frequency Check

Recommended Model/
Part Numbers
Agilent E8257D with Opt 520
Agilent 11667B
Agilent E4418B or E4419B
Agilent E4413A
Agilent 8120-4948
Agilent E2655-83202
Agilent 54855-67604
35