To Setup The Bios - Agilent Technologies Infiniium 90000 A Service Manual

Oscilloscopes
Table of Contents

Advertisement

To setup the BIOS

If the BIOS settings become corrupt, the Infiniium oscilloscope PC motherboard will not
recognize the hard drive and the oscilloscope may not start. To configure the motherboard BIOS
parameters to the default settings, follow these steps:
Connect the power cable to the Infiniium oscilloscope.
1
Connect the external keyboard to the rear panel.
2
Press Delete when the following prompt appears on the bottom of the screen:
3
Press TAB to Show POST screen, DEL to enter SETUP, F12 to select boot device.
If the prompt does not appear, or the oscilloscope does not appear to be functioning, check the
ribbon cable connectors. Otherwise, continue with the next step.
Go to Load 15" LCD Defaults and press Enter for optimized values, or go to Load
4
8.4" LCD Defaults and press Enter for fail-safe values. Select Y to load the defaults,
then press Enter.
Go to the Save and Exit Setup option and press Enter to save and exit the setup.
5
Select Y to save and exit, then press Enter.
Chapter 4: Troubleshooting

To setup the BIOS

79

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents