Chapter 4: Troubleshooting

AutoProbe Board Trouble Isolation

AutoProbe Board Trouble Isolation

AutoProbe Board Trouble Isolation
Inspect the cable between the
AutoProbe assembly and the power
board.
Cable
No
Replace bad cable.
OK?
Yes
Turn unit on.
Does
Replace
unit turn
No
AutoProbe
ON?
assembly.
Yes
Go to 'Primary
Trouble Isolation.