Performance Test Record

Agilent Technologies
Model Number _____________________
Serial Number ___________________________
Recommended Test Interval - 1 Year/2000 hours
Recommended next test date ___________________
Offset Performance Test
Zero Error Test
Vertical
Test Limits
Sensitivity
5 mV/div
-1.4 mV to +1.4 mV
10 mV/div
-1.8 mV to +1.8 mV
20 mV/div
-2.6 mV to +2.6 mV
50 mV/div
-5.0 mV to +5.0 mV
100 mV/div
-9.0 mV to +9.0 mV
200 mV/div
-17.0 mV to +17.0 mV
500 mV/div
-41.0 mV to +41.0 mV
1 V/div
-81.0 mV to +81.0 mV
Agilent 90000 Series Oscilloscopes
Tested by___________________
Work Order No.___________________
Date___________________
Ambient temperature ___________________
Channel 1
Channel 2
Channel 3
Channel 4