3 Operation
3.
Select with the Right or Left keys CAL-HOLD-TIMER, and
press the Enter key to access this function menu:
The calibration hold time is set on the first row, while the sample hold
time is set on the second row. To select one or the other, use the Right or
Left keys. To modify the time of either timer, use the Up or Down keys.
The time is in the minutes.
3.3.11

Analog 4 to 20 mA Output Calibration

This function will let the operator calibrate the 4 to 20 mA analog output to
match the display reading. A DMM configure as a DC ammeter is needed. The
DMM should be connected across the output terminals of the 4 to 20 mA output
to monitor the output current. To enter the 4 to 20 mA output adjust function,
you must:
1.
Press the System key to start the System function:
2.
Using the Right or Left arrow keys, select MORE and press
Enter. The second System screen appears:
3.
Using the Right or the Left arrow keys, select MORE and
press Enter. The third System screen appears:
3-14
Part I
Teledyne Analytical Instruments
Model 6000A