Table 4-2: Range ID Output Voltage

Range
Voltage (V)
LO
MED
HI
CAL (0-25%) 1.00
IMPORTANT: In the event of loss of flow through the analyzer, if
the vent is vented to a location of high oxygen
content, oxygen will back diffuse through the vent
line and enter the cell. For the standard B-2CXL
Micro-fuel Cell this can saturate the cell with
oxygen and require a long purge down time before
the sensor can recover. This is less of a problem
when using the Insta-Trace sensor which is
specifically designed to allow rapid recovery from
accidental high oxygen exposures.
In the event that flow is to be interrupted into the
analyzer, it is suggested that the user do one of the
following:
1. Bag the sensor in nitrogen during this time
2. Install a shut off valve on the vent port of the
3. Retrofit the instrument to use the Insta-Trace

4.10 The IP Address Function

This function is only available on web enabled instruments
(3000TA-XL-EU-I) and is used to set the IP address of the embedded
Ethernet web server coprocessor. To set the IP address, enter the system
menu:
Teledyne Analytical Instruments
Current (mA)
0.25
0.50
0.75
analyzer or somewhere within the users
sample system.
Micro-fuel Cell (see Section 3.4)
Operation
8
12
16
20