Trace Oxygen Analyzer
Section V - Emergency and First Aid Procedures
Eye Contact: Flush eyes with water for at least 15 minutes
Skin Contact: Wash affected area with plenty of water and
Ingestion: Give plenty of cold water. Do not induce
Inhalation: Liquid inhalation is unlikely.
Section VI - Handling Information
NOTE:
The oxygen sensors are sealed, and under normal circumstances, the
contents of the sensors do not present a health hazard. The following
information is given as a guide in the event that a cell leaks.
Protective clothing: Rubber gloves, chemical splash goggles.
Clean-up procedures: Wipe down the area several times with a wet
Protective measures
during cell replacement:
Disposal: Should be in accordance with all applicable
NOTE:
The above information is derived from the MSDS provided by the
manufacturer. The information is believed to be correct but does not
purport to be all inclusive and shall be used only as a guide.
Teledyne Analytical Instruments shall not be held liable for any
damage resulting from handling or from contact with the above
product.
and get immediate medical attention.
remove contaminated clothing. If burning
persists, seek medical attention.
vomiting. Seek medical attention. Do not
administer liquids to an unconscious person.
paper towel. Use a fresh towel each time.
Before opening the bag containing the sensor
cell, check the sensor cell for leakage. If the
sensor cell leaks, do not open the bag. If there
is liquid around the cell while in the
instrument, put on gloves and eye protection
before removing the cell.
state, local and federal regulations.