5.3 Description of the connection procedure
CAUTION
Opening the connection panel cover
• Only service may open the connection panel cover.
• The screw torque may not exceed 0.75 Nm +/- 0.05 Nm.
• Patients must not be present when the cover is open.
CAUTION
Connector
Connectors may only be plugged in or removed by Service when the device is switched
off.
Prerequisite
The monitor connection panel is freely accessible and the monitor is in a stable installation
position.
5.3.1 Connect the DVI receiver modules
If you want to use the TDL transmission links from EIZO with the monitor, connect the corre-
sponding DVI receiver modules as follows:
1. Loosen the connector panel Combi-Torx screws and remove the cover.
2. Loosen the screws on the DVI module holders and remove the holders.
CuratOR LX550W
Instructions for Use, 04/2021
5.3 Description of the connection procedure
Screws for the DVI module mounts
Connecting
DVI module mounts
23