Connecting

5.3 Description of connection procedure

Grounding screw
The additional protective conductor is connected to the grounding screw.
Service connection
The service connection is used by Service for software updates.
Serial interface for photometer
The serial interface is freely accessible under the right fan.
NOTICE
Connecting a photometer
• Only service can connect or disconnect a photometer.
• Only photometers tested for calibrating the monitor may be connected to the serial inter-
face.
• A photometer must not be connected in the presence of the patient.
5.3 Description of connection procedure
CAUTION
Opening the connector panel cover
Only service may open the connector panel cover. Patients must not be present when the
cover is open.
CAUTION
Connector
Connectors may only be plugged in or removed by Service when the device is switched
off.
Prerequisite
The monitor must be installed correctly.
Connecting
1. Loosen the connector panel Combi-Torx screws.
2. Remove the connector panel cover.
3. Connect the appliance plug to the monitor power socket.
4. Secure appliance cable with cable ties to the lug to prevent unintentional loosening
(marked with a red circle).
22
RadiForce LS580W
Instruction for Use, 02/2014