DO NOT use this procedure on an Agilent 83485A or other optical module. It
C A U T I O N
will destroy the factory optical/electrical calibration factors.
If all of the above steps check out okay, and the plug-in module still does not
operate properly, then the problem is beyond the scope of this book. Refer to
the optional
54752A,B Service Guide,
department.
The calibration status of the plug-in modules is listed near the bottom
of the display, in the box labeled \
of the plug-in module is listed next to the appropriate slot number,
then the mainframe has identi ed the plug-in.
If \
" is displayed instead of the model number of the plug-in
~known
module, remove and reinsert the plug-in module in the same slot. If
\
" is still displayed, then the memory contents of the plug-in
~known
module are corrupt.
a. To reinitialize the plug-in memory, press:
Utility
4
5
NNNNNNNNNNNNNNNNNNNNNNN
Service
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Plug-in . . .
b. Select the appropriate plug-in by selecting slot 1 and 2, or 3 and 4
by pressing:
NNNNNNNNNNNNNNNNNNNNNNN
1 or 3
Plug-in
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Initialize plug-in memory
Agilent 83483A, 54751A Service Guide
or return the plug-in module to a quali ed service
FINAL TRIM SIZE : 7.5 in x 9.0 in
In Case of Diculty
If the plug-in does not operate
". If the model number
Plug-ins
Agilent 83484A,B and
or
5-5