Speci cations and Characteristics
Speci cations
Agilent 83484A,B and Agilent 54752A,B Vertical Speci cations
Bandwidth ( 3 dB)
dc Accuracy|single marker
1
26.5 GHz bandwidth
50 GHz bandwidth
dc Di erence|two marker accuracy on same channel
26.5 GHz bandwidth
50 GHz bandwidth
Transition Time (10%{90%) characteristic,
calculated from T=0.35/BW, electrical
26.5 GHz bandwidth
50 GHz bandwidth
RMS Noise
Typical
26.5 GHz
50 GHz
Maximum
26.5 GHz
50 GHz
Scale Factor
Minimum
Maximum
dc O set Range
Nominal Input Impedance
It is recommended that a user vertical calibration be performed after every 10 hours of continuous use or if the temperature has changed by greater than 2
1
from the previous vertical calibration.
Where 1T
represents the temperature change in Celsius from the last user vertical calibration. Note that the temperature term goes to zero upon
2
cal
execution of a vertical calibration.
4-6
dc to 26.5 or 50 GHz, user selectable
6
0.4% of full scale
6
6
2 mV
1.2% (reading channel o set)
6
(2%/
C) (1T
) (reading)
2

cal
6
0.4% of full scale
6
6
2 mV
2% (reading channel o set)
6
(2%/
C) (1T
) (reading)

cal
1
6
0.8% of full scale
6
1.2% of delta reading
6
(2%/
C) (1T
) (reading)
2

cal
6
0.8% of full scale
6
2% of delta reading
6
(2%/
C) (1T
) (delta reading)

cal
13.2 ps
7.0 ps
0.46 mV
0.92 mV
0.75 mV
1.5 mV
full scale is eight divisions
1 mV/div
100 mV/div
6
500 mV
50
FINAL TRIM SIZE : 7.5 in x 9.0 in
2
2
C