The Scale softkey controls the vertical scaling of the waveform. If the ne
mode is o , then the knob and arrow keys change the vertical scaling in a
1-2-5 sequence. When ne mode is on, the knob and arrow keys change the
vertical scaling in 1 mV increments. You can also use the keypad to enter
values in 1 mV increments, independent of the ne mode selection.
The units the scale is displayed in depend on the unit of measure selected
with the Units softkey. The choices for units are volts, watts, amperes, or
unknown.
Key Path
4
aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
The O set softkey moves the waveform vertically. It is similar to the position
control on analog oscilloscopes. The advantage of digital o set is that it is
calibrated. The o set voltage is the voltage at the center of the graticule
area, and the range of o set is 500 mV. You can use the knob, arrow keys, or
keypad to change the o set setting. The ne mode also works with o set.
When an Agilent 54700-series active probe is used with the plug-in module
and is connected to the probe power connector adjacent to the channel input,
the o set control adjusts the external scale factor and o set of the hybrid
inside the active probe. A probe connected to the auxiliary power connector
adjacent to the trigger input will function, but the channel scale factor will
not be adjusted automatically.
Key Path
4

Scale

NNNNNNNNNNNNNNNNN

Scale
Channel
5

Offset

NNNNNNNNNNNNNNNNNNNN

Offset
Channel
5