Channel Setup Menu

Bandwidth. . .

aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
Agilent 83483A, 54751A only
You can use the Bandwidth function to select either the 12.4 GHz or the
20 GHz bandwidth.
Agilent 83484A,B 54752A,B only
You can use the Bandwidth function to select either the 26.5 GHz or the
50 GHz bandwidth.
Key Path
4
Channel

Channel autoscale

aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
The Channel Autoscale function provides a convenient and fast method for
determining the standard vertical scale setting with the highest resolution
that will not clip the waveform. Timebase and trigger settings are not
a ected.
This function is useful in manufacturing environments where the timebase
and trigger settings remain constant and only the vertical scale needs to be
adjusted for signal level variations in multiple DUTs.