Calibration Overview
Factory Mainframe Calibrations
The temperature change is displayed at the top of the display as shown in the
following gure.
Figure 3-1. Current Frame 1Temp condition
6
If the
listing is greater than
5
C, then the
Current Frame 1Temp
mainframe should either be calibrated at the current operating temperature or
be placed in an ambient air temperature that is within 5
C of the temperature
of the current calibration.
3-5
FINAL TRIM SIZE : 7.5 in x 9.0 in