Agilent Technologies 54751A Manual page 36

Plug-in modules
Hide thumbs Also See for 54751A:
Table of Contents

Advertisement

Channel Setup Menu
The instrument displays
whether the last plug-in module calibration is still valid. A calibration can be
invalidated if:

The mainframe has cycled power.

The plug-in has been repaired, reprogrammed, or removed from the
mainframe.

The instrument's operating temperature has changed and remains more
than 5
C from the temperature at which the Plug-in calibration was

performed.
indicates the plug-in module vertical calibration is invalid.
Uncalibrated
The Plug-in function lists the model number, serial number, date, time, and
temperature delta. The temperature 1 is the temperature change from the
temperature of the mainframe when the last calibration was performed. If
this temperature 1 is greater than
then you must perform a plug-in module calibration to achieve the speci ed
dc accuracy.
Connect a voltage probe to the plug-in and then press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Calibrate probe
The instrument calibrates to the tip of the probe by setting the probe
attenuation to the actual attenuation ratio of the probe. The instrument
also automatically compensates for any o set that the probe may introduce.
The CAL signal is internally routed to the probe tip for Agilent probes.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Calibrate
4
Channel
5
2-12
FINAL TRIM SIZE : 7.5 in x 9.0 in
or
Calibrated
Uncalibrated
6
5
C since the last mainframe calibration,

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Calibrate probe
, depending on