Bard CritiCore 000002 Service Manual page 43

Advertisement

Removing the Front
Equipment Required
Enclosure Assembly
e
Small standard screwdriver
«
No. 2 Phillips screwdriver
*
Needle nose pliers
Caution:
The Bard CritiCore Monitor main and display PCB's contain
AN
static sensitive electronic devices.
Do not attempt to repair any
internal componenis without proper ESD (Electro-Static
Discharge) handling equipment.
Failure to properly ground both
your body and your work station could result in damage to the
Monitor.
Procedure
Note:
Review the contents of the front enclosure replacement parts kit
(see the Replacement Parts List on page 64}, to confirm which
parts you may need to keep/discard during disassembly.
Remove the hooks per the Removing the Hooks procedure on page 36.
2.
Remove the rear enclosure per the Removing the Rear Enclosure
procedure on page 39.
3.
Remove the two (2) 6-32x3/8" Phillips pan-head screws that secure the
transducer, and pull out the transducer and transducer cap from the front
enclosure.as illustrated in Figure 54.
4.
Remove the four (4) 6-32x3/16" Phillips pan-head screws which secure
the main PCB to the front enclosure (see Figure 55). Lift up and remove
the main PCB.
Installing the Front
Equipment Required
Enclosure Assembly
e
Small standard screwdriver
e
No. 2 Phillips screwdriver
e
Needie позе pliers
Caution:
The Bard CritiCore Monitor main and display PCB's contain
static sensitive electronic devices.
Do not attempt to repair any
internal components without proper ESD (Electro-Static
Discharge) handling equipment.
Failure to properly ground both
your body and your work station could result in damage to the
Monitor.
Procedure
1.
Install the four (4) 6-32x3/16" Phillips pan-head screws and attach the
main PCB to the front enclosure (see Figure 55).
Note:
Ensure that there is no dirt, oil or contamination on the
transducer cap or transducer.
2.
Place the transducer cap onto the transducer.
Install the two (2) 6-
32x3/8" Phillips pan-head screws and attach the transducer/cap to the
front enctosure as iftustrated in Figure 54. Ensure that the transducer
cable will not be pinched when the rear enclosure is attached.
; Figure 55
Bard CritiCore Monitor Service Manual
Page 37