Procedure
1.
Install the bezel enclosure to the bezel top assembly by installing the
four (4) 6-32x1/4" Phillips pan-head screws.
2.
Reconnect the display cable connector to the main PCB (see Figure 12).
Ensure that the cable has been fully inserted and that the ejector latches
are firmly closed.
3.
Place the rear enclosure onto the front enclosure, ensuring that the
plastic blank insert is held into position as illustrated in Figure 11. Instait
the rear enclosure by installing the four (4) 6-32x3/8" Phillips pan-head
screws (see Figures 9 and 10).
Figure 11
4.
Ensure that the display assembly is correctly oriented so that the holes
for the hinge pins are clear.
5.
Insert the two hinge pins (see Figure 8), and replace the snap rings.
6.
Insert and connect the battery packs and install the battery doors.
7.
Stand the Monitor up, open the rear control panel door and reattach the
SET VALUE and TEST SWITCH rotary knobs, ensuring that the arrows
on the two knobs are oriented in the same direction that they were in
when they were removed.
8.
Perform the Functional Checkout procedure (see page 7) to ensure that
alt display and general Monitor functions are working properly.
Bezel Top
The bezel top is the outermost plastic enclosure of the display assembly.
The membrane switch, bezel enclosure and display PCB/LCD assembly aré
directly attached to the bezel top. The bezel top is assembled to the bezel
enclosure which is hinged by the hinge pins to the front enclosure. Figure 13
shows a bezel top with the membrane switch installed.
Figure 13
Testing the Bezel Top
Visually inspect the bezel top for damage.
If the bezel top is broken or
cracked, it may allow fluids to enter into the display PCB and should be
replaced.
Removing the Bezel Top
Equipment Required
No. 1 Phillips screwdriver
+
No. 2 Phillips screwdriver
*
Needle nose pliers
static sensitive electronic devices.
Do not attempt to repair any
internal components without proper ESD (Electro-Static
Discharge) handling equipment.
Failure to properly ground both
your body and your work station could result in damage to the
Monitor.