Removing the Hooks
Equipment Required
*
Smali standard screwdriver
Procedure
Note:
Review the contents of the Hook replacement parts kit (see the
Replacement Parts List on page 64), to confirm which parts you
may need to keep/discard during disassembly.
1.
Separate the hook caps by inserting a small jeweler's screwdriver
between each hook cap and the arm as illustrated in Figure 51.
Figure 51
Figure 52
2.
Using a small jeweler's screwdriver, push the dowel pin out of the arm
insert and remove the hook as illustrated in Figure 52.
Installing the Hooks
Procedure
1.
Replace the hook onto the arm by inserting the dowel pin through the
hook and arm insert. Ensure that the hook is installed with the proper
orientation: With the Monitor facing you and the arms hanging down, the
hook(s) opening should face towards the front of the Monitor.
2.
Secure-the dowel pin by replacing the hook caps onto the arm.
3.
Visually inspect the hook to ensure that the hook caps are secure and
that the hook moves freely.
Front Enclosure Assembly
The front enclosure assembly contains the front enclosure, handle, arms,
and select hardware. Attached to the front enclosure are the plastic retainers
and attraction plates.
The plastic retainers are located on the inside of the
front of the Monitor.
These retainers allow for the disposable container to
remain seated in the Monitor, and atop the transducer.
The attraction plates
are adhered to the upper portion of the Monitor next to the arms. The
magnets on the bezel enclosure are drawn to the attraction plates. The force
of attraction is necessary to apply pressure to the top of the disposable
container. The force on the container cap ensures that the container is firmly
seated' on the transducer cap.
This in turn allows for minimal ultrasound
signal loss when the container has little or no fluid inside.
Directly mounied to the front enclosure are the display assembly, main PCB,
transducer/transducer cap and rear enclosure.
Note:
Figure 53 illustrates the front enclosure without the handle and
arms.
Figure 53
-
Testing the Front Enclosure
—_Vsuaily inspect the front enclosure, if the front enclosure is cracked or
Assembly
broken it must be replaced.