5.
Repeat step 4 after each minute increment for 10 consecutive minutes,
adding only a small amount of water each time.
After exactly 10 minutes the current Flow Rate display should change
from dashes *---" to a value that is 6 times the value in the Present
Interval Output display.
Not Level Indicator
With the Monitor ON, tilt the Monitor.
Verify that the Not Level status indicator becomes lit.
3.
Return the Monitor to a level position and verify that the Not Level
indicator disappears.
Full Container Alarm
1.
Open the drain clamp on the upper container or use an irrigation syringe
to fill the container in the Monitor to 2050 mi of water.
2.
Verify that the Full Container status indicator flashes and an audible
alarm is heard.
Note that the BEEPER toggle switch should be in the
AUDIBLE position.
3.
Press the Container Volume switch and verify that the Container Volume
displayed is greater than 2000mi and is flashing.
4.
Drain about 200ml from the Monitor container and verify that the Full
Container status indicator is not lit and the audible alarm stops beeping.
5.
Press the. Container Volume switch and verify that the Container Volume
displayed is less than 2000ml and is not flashing.
High Temperature Alarm
Before performing this procedure, remove any temperature probes or
calibration plugs and reset the Monitor.
'
1.
Set the TEMP ALARM toggle switch to the ENABLED position and the
BEEPER toggle switch to the AUDIBLE position.
Verify that the bell icon appears in the upper right corer of the Bladder
Temperature display.
2.
Turn the SET VALUE rotary switch to the HIGH TEMP. ALARM position.
3.
Verify that the Set Value status indicator and 101.0°F flashes on the
display. The High Alarm and Check Probe status indicators should also
be lit.
Note:
The 101.0°F will appear only if the temperature alarm default
limit has not been reset or if both battery packs have been
removed for an extended period of time.
4.
Use the Tl switches on the membrane switch to set the alarm limit to
98.4°F.
5.
Turn the SET VALUE rotary switch to the NORMAL OPERATION
position.
Insert a 98.6°F calibration plug. Verify that the High Alarm
status indicator is lit, the value in the Bladder Temperature display is
98.6+ 0.2 °F,:and is flashing and the audible alarm is beeping.
Page 10
Bard CritiCore Monitor Service Manual