Low Temperature Alarm
Backlighting
Completion
6.
Press the Alarm silence switch to verify that the audible alarm stops, the
value in the Bladder Temperature display continues to flash and the
Alarm Silenced status indicator is lit.
7.
Reset the High Temperature Alarm limit to the preferred high
temperature alarm value.
Remove the calibration plug and reset the Monitor.
2.
Turn the SET VALUE rotary switch to the LOW TEMP. ALARM position
and verify that the Set Value status indicator and 50.0°F (10.0°C) flashes
on the display. The Low Alarm and Check Probe status indicators
should also be lit.
Note:
The 50°F will appear only if the temperature alarm default limit
has not been reset or if both battery packs have been removed
for an extended period of time.
3.
Use the TL switches on ihe membrane switch to set the alarm limit to
99.0°F.
4.
Tum the SET VALUE rotary switch to the NORMAL OPERATION
position.
-
5.
insert a 98.6°F calibration plug. Verify that the Low Alarm status
indicator is lit and the audible alarm is beeping.
6.
Reset the Low Temperature Alarm limit to the preferred low temperature
alarm value.
Place the BACKLIGHT toggle switch in the ON position.
Press any switch on the membrane switch.
3.
Do not touch any other key. Verify that the display backlight is on and
that it will automatically shut off after 30 seconds.
Note:
It may be necessary to dim the room lighting before the Monitor
backlight is visible.
4.
If the Monitor will be used in well-lit conditions retum the BACKLIGHT
toggle switch to the OFF position to maximize battery life.
At this point the Functional Checkout Procedure is complete.
Store the two
containers used to hold fluid so they can be used to check the Monitor as
required.
Bard CritiCore Monitor Service Manual