Functional
Chec
t
Product
Identification
Self Test
The Functional Checkout procedure is designed to verify that the CritiCore
Monitor is functioning properly before it is placed into service. This
procedure is available in the Operator's Manual, which is supplied with every
Monitor purchased.
The Operator's Manual can also be purchased
separately (see the Replacement Parts List on page 68). The Functional
Checkout procedure is repeated here for reference.
In addition to the
Functional Checkout procedure are the use of the T, - T, Monitor test modes.
The use of these test modes will be discussed in the Additional Testing
Modes section on page 12.
1.
the Bard CritiCore Monitor has been turned on recently, be sure it has
been turned off for at least 10 minutes before initiating the self test.
2.
Hang the Monitor from a stable levet bracket or set it on a stable level
surface.
3.
Clean both the cavity that holds the collection container and the silicone
rubber transducer cap located at the bottom of the cavity.
Note:
Use a soft non-abrasive cloth dampened with warm water and
mild detergent, alcohol or a non-staining disinfectant.
4.
Open the rear control panel door to reveal the rear panel switches
5.
Setthe toggle and rotary switches as follows:
Set Format
Set Value
Test Switch
CLOCK
42 Hr.
Normal Operation
Normal Operation
TEMPERATURE
°F
BACKLIGHT
Of
FLOW RATE DISPLAY
Off
FLOW RATE ALARM
Disabled
TEMP. ALARM
Disabled
BEEPER
Audible
6.
Press the ON/OFF switch on the membrane switch to turn the Monitor
ON.
The Monitor will emit a short "beep" and perform a self test. This self test
should take about 4 seconds.
After the self test the Monitor will
automatically enter the normal operation mode.
Note:
If the Monitor fails the self test, an error code "PPPP" will appear
in the top display.
Once the self test is completed and successful, the display should
resemble Figure 4 with the exception of Not Level Indicator.
Note:
The current time of day, time of the previously selected interval
and interval length may differ from Figure 4.