Testing the Temperature
Equipment Required
Jack Connector
*
Calibration piug or temperature-sensing Foley catheter
Procedure
1.
Turn the Monitor on.
2.
Connect a calibration plug or temperature-sensing Foley catheter to:the
jack on the display assembly.
Note that there should be a temperature
reading in the Bladder Temperature window of the display (The readings
will vary depending on the air temperature as well as the F° or © setting
on the Monitor).
3.
Verify that there are no intermittent readings by gently moving the probe,
and pulling the probe out and plugging it back in several times.
If
intermittent readings are observed during this test, the temperature jack
must be repiaced.
Note:
The calibration of Bladder Temperature is not performed by
adjusting any components on the display PCB.
If you suspect
your Monitor is out of temperature calibration, see the Installing
and Calibrating the Main PCB procedure (on page 58) for
information.
:
Removing the Temperature
Equipment Required
Jack Connector
e
'No.1 Phillips screwdriver
+
No. 2 Phillips screwdriver
*
Small standard screwdriver
+
Needle nose pliers
e
Desoldering tools
Caution:
The Bard CritiCore Monitor main and display PCB's contain
static sensitive electronic devices.
Do not attempt to repair any
internal components without proper ESD (Electro-Static
Discharge) handling equipment.
Failure to properly ground both
your body and your work station could result in damage to the
Monitor.
Note:
Review the contents of the temperature jack replacement parts
kit (see the Replacement Parts List on page 64), to confirm
which parts you may need to keep/discard during disassembly.
Figure 23
Procedure
1.
Remove the bezel enclosure per the Removing the Bezel Enclosure
procedure on page 20.
2..
Disconnect the membrane switch mylar ribbon:cable from the display
PCB (see Figure 23).
3.
Remove the four assembly screws which attach. the display. PCB/LCD.
assembly to the bezel top.
Separate the display PCB/LCD-assembly.
from the bezel top by gently pulling up on the display: PCB comers as.
iltustrated in Figure 24.