Check Container Status
1.
Lift the display assembly and install an empty disposable container into
the Monitors front cavity.
`
2.
Lower the iid ensuring that it completely closes and traps the container
within the Monitor.
3.
Verify that the Check Container status indicator on the display is not lit,
and "O ML" appears in the Present Interval Output display.
Temperature
1.
Install a temperature-sensing Foley catheter or a temperature calibration
plug into the temperature jack on the left side of the display assembly.
Note:
A precision (0.1% or better) decade box set to 1954.70 and
connected to a 3.6mm mono mini-phone plug may be used
instead of a calibration plug. See page 59 for table of calibration
plug resistances.
2.
Verify that the Check Probe status indicator disappears and that a
temperature reading appears in the Bladder Temperature display.
If
employing a temperature calibration plug the value should be 98.6°F +
02Ε.
3.
Remove the temperature calibration plug or temperature-sensing Foley
catheter and verify that the Bladder Temperature window displays "--.-"
and that the Check Probe status indicator is lit.
System Clock
1.
Tur the SET VALUE rotary switch on the rear control panel to the
HOUR position. Verify that the hour segment, PM/AM and Set Value
status indicators are flashing.
2.
Use the Tl switches on the membrane switch to set the system clock to
the correct hour..
3.
Turn the SET VALUE rotary switch on the rear control panel to the
MINUTE position. Verify that the minute segment and Set Value status
indicator are flashing.
4.
Use the T switches on the membrane switch to set the minutes of the
clock to the correct time of day.
5.
Return the SET VALUE rotary switch to the NORMAL OPERATION
position and verify that the correct time appears on the display. Verify
that the Set Value status indicator is not lit.
Intervals
1.
Turn the SET VALUE rotary switch to the INTERVAL position. Verify
that the interval segment, HA/MIN and Set Value status indicators are
flashing.
2.
Use the Tl switches onthe membrane switch to display the following
intervals: 5 MIN; 10 MIN,
15 MIN, 20 MIN; 30 MIN; 1HR; 2HR.
3.
Set the interval length to the preferred interval and return the SET
VALUE rotary switch to the NORMAL OPERATION position.
Page 8