Volume Measurement Troubleshooting Flow Chart
Monitor is not measuring volume
consistently or correctly.
Test the STRTXM charge
4
signal (page 51).
4
Verify the following:
The container holds less than
2000mi of fluid.
"Not Level" indicator is not lit.
"Check Container" indicator is not lit.
Transducer cap is clean.
Does STRTXM signa
check out OK?
N
Replace Main Circuit
o:
Board.
Yes
y
Test the Transducer signal
(page 50).
_ Test the Display LCD (page 29).
Does Transducer
?
Does LCD check out OK?
signal check out OK?
No
Replace Transducer.
Yes
No
Replace LCD.
Does problem persist?
ο
Problem solved.
Yes
Send unit to Bard for service.
Bard CritiCore Monitor Service Manual