Examples Of Programming - HP 5340A Operating And Service Manual

Frequency counter
Hide thumbs Also See for 5340A:
Table of Contents

Advertisement

Model 5340A
Installation and Remote Programming
(3) Clears the display.
(4) Initializes the phase locked loops forcing
(5) Starts measurement phase of the 5340A's operating cycle.
c.
Sample Triggers Instruction (I):
(1) Can be given only if the 5340A
operating cycle. If given
(2) Does not change the information in the program storage cells.
(3) Does not clear the display.
(4) Does not initialize the phase locked loops, i.e., the 5340A does not go through a
search procedure unless it h a s lost phase lock since the previous measurement.
(5) Starts the measurement phase of the 5340A's operating cycle.

2-55. Examples of Programming

2-56. Assume that it is desired to program a 5340A for a measurment of approximately 3.5 GHz
to a
subsequently output when so instructed. One method of programming this is shown in Table 2-6.
(Assume the 5340A listen address is ASCII
2-57. Another example
of
desired to program a 5340A using a 3260A Mark Sense Programmer for a n automatic measure-
ment to a
1
program card.
,
2-18
is stopped in the Sample Rate phase of its
at any other time it will be ignored by 5340A.
and the talk address is ASCII
programming is with a mark sense card reader. Assume that it is
,
a new search procedure.
"C".)