2-1. INTRODUCTION

2-2. This section tells how to set up the 5340A Frequency Counter. Instructions for unpacking,
inspecting, installing, and remote programming are included.

2-3. UNPACKING AND INSPECTION

2-4. If the shipping carton is damaged, inspect the counter for visible damage (scratches, dents,
etc.). If the counter is damaged, notify the carrier and the nearest Hewlett-Packard Sales and
Service Office immediately (offices are listed at the back of this manual). Keep the shipping
carton and packing material for the carrier's inspection. The Hewlett-Packard Sales and Service
Office will arrange for repair or replacement of your instrument without waiting for the claim
against the. camer to be settled.

2-5. INSTALLATION REQUIREMENTS

3
. -
2-6. LINE VOLTAGE REQUIREMENTS. The 5340A is equipped with a line voltage switch to
select 115-volt or 230-volt ac operation. Before applying power, the rear panel screwdriver-
operated switch must be set to the correct position ("115" or "230" visible) and the correct fuse
(as
labeled on the rear panel) must be installed. See Figure 3-4 for rear panel features.
2-7.
LINE FREQUENCY REQUIREMENTS.
between 48 Hz and 66
THREE CONDUCTOR POWER CABLE.
2-8.
grounded three-conductor detachable power cable. The male connector end is a NEMA type
connector, and the female connector end is a C.E.E. type connector that mates with the 5340A
rear panel power connector.
NEMA grounded third conductor. If the line power receptacle is a standard two-pin type instead
of the NEMA three-pin receptacle, use a two-to-three pin adaptor (HP Part No. 8120-1348) and
connect the green pigtail on the adaptor to ground.
2-9. TEMPERATURE LIMITS. Maximum and minimum allowable operating temperatures ake
listed in Table 1-3. If these limits are exceeded at the installation site, auxiliary cooling or
heating should be used to keep the environment within limits.
2-10. RACK INSTALLATION. The counter is ready for bench operation as shipped from the
factory. Additional parts necessary for rack mounting are packaged with the instrument. To
convert the instrument to 'rack installation, refer to Figure 6-1 for parts identification and pro-
ceed as follows:
a. Remove tilt stand MP14 by removing the two outside front feet M P l l from the bottom
cover MP9.
the foot toward the center of the instrument.
b. Remove the remaining three feet from the bottom cover.
SECTION II
INSTALLATION AND REMOTE PROGRAMMING
BEFORE CONNECTING THE INSTRUMENT TO AC
POWER LINES, BE SURE THAT THE LINE SELECTOR
IS PROPERLY POSITIONED.
Hz.
Connect the power cable to a power source receptacle with a
The feet are removed by pressing the foot-release button and sliding
Installation and Remote Programming
The counter will operate at line frequencies
To protect the operator, the counter uses a
Model 5340A