Philips VueLink M1032A Service Booklet page 76

Table of Contents

Advertisement

Datex Capnomac (II and Ultima) Gas Analyzer
76
Supported Devices:
Connection:
Switch Settings:
l = language
Datex Capnomac II
Datex Capnomac II four channel analog
output option
Datex Ultima
All units shipped after Feb 21st 1994
Analog / RS-232 Digital combination
Baud Rate:
1200
Word Length:
8 bits
Stop Bits:
1
Parity:
None
Analog:
1 wave on Capnomac II
3 waves on Capnomac II
four channel analog output option
4 waves on Capnomac
Ultima
Factory default
Gas Analyzers