The configurations for each numeric are provided in the following table.
Channel
Label
Wave 1
Pleth
Wave 2
CO2
Numeric 1
SpO2
Numeric 2
EtCO2
Numeric 3
Resp
Numeric 4
EtN2O
Numeric 5
Pulse
The low and high grid values have to meet the restrictions outlined in the VueLink
Handbook.
Nellcor N1000 and N2500 Multi-Function Monitor
Value Range =
Unit
Voltage Range
-
0 to 100 = 0 to 10 V
mmHg
0 to 76 mmHg = 0 to 10 V
%
0 to 100% = 0 to 10 V
mmHg
0 to 76 mmHg = 0 to 10 V
rpm
0 to 75 rpm = 0 to 10 V
mmHg
0 to 760 mmHg = 0 to 10 V
rpm
0 to 250 bpm = 0 to 10 V
Decimal
Represent.
x
x
x
x
x
x