Datex Capnomac (II and Ultima) Gas Analyzer

Start Up Procedure
Note
Warning
Gas Analyzers
The driver automatically identifies the connected device at start-up. For this reason,
the following start-up procedure must be followed whenever the user changes the
Datex Device, even if it is the same model:
1. Turn off the Datex Device.
2. Plug in and Set up the VueLink Module and connect the cable.
3. Turn on the Datex Device.
To ensure proper operation of the VueLink module and accurate data, the start up
procedure described above must be carried out:-
If the Datex device is turned off.
If the VueLink module or cable is unplugged.
Whenever the Datex device is selected.
After re-entering monitoring mode.
In order to avoid damaging the Datex device, the VueLink cable should
only be connected to, or disconnected from, the Datex device when it is
switched off.
Device Driver Name:
Device Driver P/N:
Datex Capnomac (II and Ultima) Gas Analyzer
DATEX
M1032-B13rl
where:
r = revision
75