Waves are only available directly from the device (Narkomed 2C or Vitalert
3200); therefore, connect VueLink to the device whose waves are desired. (The
Communications Hub does not provide waves.)
If the patient monitor RS-232 MECIF interface is being used, then configure
MECIF only for the highest (in letter, e.g., C is higher than B) used port on the
Communications Hub or Vitalert 3200, or only to Port B on the Narkomed 2C.
Otherwise, incorrect data may be displayed.
If all of the following conditions are true:
The patient monitor RS-232 MECIF interface is connected to the Narkomed
2C or Communications Hub.
Another Type B VueLink module (Ventilator, Gas Analyzer, Anesthesia
Machine or Auxiliary Plus) is being used.
The above VueLink module is measuring one or more numerics which are also
output by the Narkomed VueLink interface. (For example, the Philips M1025A/
B Anesthetic Gas Analyzer numerics include !imCO2, !etCO2, !inN2O, etN2O,
!FIO2, !inAGT, !etAGT, !PULSE, !SpO2 and !AWRR. These correspond to the
Narkomed VueLink interface numerics: IMCO2, ETCO2, inN2O, etN2O,
FIO2, inAGT, etAGT, PULSE, SpO2 and AWRR.)
Then the following must be performed to avoid the display of incorrect data:
Each duplicate numeric (from condition (c)) should NOT be selected on the
Narkomed VueLink interface. For example, if the Philips M1025A/B Anesthetic
Gas Analyzer VueLink interface is also being used, then the IMCO2, ETCO2,
inN2O, etN2O, FIO2, inAGT, etAGT, PULSE, SpO2 and AWRR numerics
must not be selected on the Narkomed VueLink interface.
Anesthesia Machines