Step 1: Ensure that all the necessary hardware is available
Note
Free Analog Device Information
(10 Minutes)
You will need:-
An M1032A #A04 VueLink Module
4m Free Analog (open end) cable M1032A #K6A or M1182A #K6A
or
2m Free Analog cable M1032A #K60 or M1182A #K60.
A connector to fit the ELSA/EAS Analog Interface connector:
ITT Cannon 25-pole, female
Standard hood with snap adapter
This connector is provided locally by the hospital or by the manufacturer; it will not
be set up at SMO/SLI. (The connector provided with the free analog cable cannot
be used).
Engström ELSA/EAS Anesthesia System