Device Driver Name:
Device Driver P/N:
Supported Devices:
Connection:
The 15-pin connector should be connected to the analog port on the ventilator. The
9-pin connector should either be connected to the RS-232 port on the ventilator, if
BEAR 1000
M1032-B28rl
where:
r = revision
l = language
Bear 1000 (Basic and Comprehensive) SW
Release 9.7
Bear GD 1000 Graphics Display (optional)
Software P/N 51001-02007
Analog (2 waves) / RS-232 Digital combination.
No graphics display:
Baud Rate:
9600 (Bear 1000 ventilator)
Graphics display:
Baud Rate:
19200 (Bear 1000 ventilator)
9600 (graphics display)
Word Length:
8-bits (fixed)
Stop Bits:
1 (fixed)
Parity:
None (fixed)