Philips VueLink M1032A Service Booklet page 180

Table of Contents

Advertisement

Nellcor N1000 and N2500 Multi-Function Monitor
Step 5: Configure each Nellcor analog output
180
(30 Minutes)
The following settings should be stored in slot 10 (defaults) of the Nellcor device:
Analog
Output Port
1
Pulse wave
2
CO
3
SaO
4
End-Tidal CO
5
Resp. Rate
6
End-Tidal N
7
Pulse Rate
Signal
wave
2
%
2
2
O
2
Full scale output voltage
10.00V
10.00V =76 mmHg
10.00V = 100%
10.00V = 76 mmHg
10.00V = 75/min
10.00V = 760 mmHg
10.00V = 250/min