Step 7: Connect the free analog cable to the TCM3 analog outputs
Step 8: Verify configuration with the users
Free Analog Device Information
Channel
Label
Numeric 1
TCPO
Numeric 2
CPCO
1. You can configure the number of decimal places by moving the decimal point, using the
left/right arrow keys. The decimal representation in the table above refers to the Value
Range. Please ensure that they are configured as shown.
(10 Minutes)
(15 Minutes)
Ensure that the correct numerics are being interfaced from the TCM3. Simulate an
alarm to ensure that the alarming capability functions correctly.
Radiometer Transcutaneous pO2/pCO2 Monitor
Value Range =
Unit
Voltage Range
mmHg
0 to 200 mmHg = 0 to 1 V
2
mmHg
0 to 200 mmHg = 0 to 2 V
2
1
Decimal
Represent.
x
x