Technical Data; General Technical Specifications; Rack-Mounted Housing - Siemens SIPROCESS GA700 Compact Operating Instructions

Continuous gas analytics
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Technical data

7.1

General technical specifications

Note
Minimum wait time before opening is 10 minutes.
7.1.1

Rack-mounted housing

Table 7-1
Rack-mounted housing: General technical data
General
Installation position
Weight
Degree of protection
Electrical characteristics
Auxiliary power supply
Power consumption
EMC interference im‐
munity (electromag‐
netic compatibility)
during operation:
Electrical safety
Equipotential bonding on housing
Transient protection if the type of protection Ex e is
used
Power unit cable
SIPROCESS GA700 rack and wall mounted devices
Compact Operating Instructions, 06/2023, A5E35134047-AB
Horizontal
Approx. 8.5 kg
IP20 according to EN / IEC 60529
with one and/or two OXY‐
MAT 7 / ULTRAMAT 7 units
with one or two CALOMAT 7
units
100 to 240 V AC (nominal range of use 85 to 264 V)
50 to 60 Hz (nominal range of use 47 to 63 Hz)
Max. 280 VA
in accordance with the standard requirements of NA‐
MUR NE21 (05/2006) and EN / IEC 61326-1
in accordance with standard requirements EN 61326-1
in accordance with EN / IEC 61010-1, overvoltage category II and in‐
stallation category II
4 mm² (11 AWG) (For installation in Ex areas)
It has to be ensured that the transient protection is set to a value 500
V for 230 V mains power and 260 V for 120 V mains power. Ensure that
voltages above 544 V can never occur.
Cable diameter: max. Ø 10 mm (0.39")
Core cross-section: max. 3 x 2.1 mm
mm² (3 x 18 AWG)
Ensure that the cross section of PE is always equal or greater than the
cross section of line and neutral.
7
(3 x 14 AWG) min. 3 x 0.75
2
87