Ex P Safety Equipment - Siemens SIPROCESS GA700 Compact Operating Instructions

Continuous gas analytics
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Note
When leaving bypass mode, the purging unit F870S returns to the same mode it was in
previously (pre-purging phase or purging phase).
B.1.1.2

Ex p safety equipment

Pneumatic and electric connections of the Ex p control system (example)
A
Ex p controller F870S
B
Ex power relay SR853
C
Ex interface relay SR852, 16-pin
D
Proportional valve
E
Pressure regulating valve for purg‐
ing gas with gauge
Power connection
Purging gas connection to the Ex p controller F870S, 12 mm (0.47") (outer diameter)
Purging gas inlet
Purging gas outlet
Purging gas connection from Ex p controller F870S, 12 mm (0.47") (outer diameter)
Purging gas connection; supply pressure 1000 hPa (rel.)
Connection from and to Ex interface relay SR852
Exhaust air outlet. If appropriate connection of a 1" line, max. length 20 m
Figure B-2
Gas path of the Ex p safety equipment, Gönnheimer Elektronik GmbH
Connection information
The supply pressure has to be sufficiently large (2.5 ... 5 bar).
SIPROCESS GA700 rack and wall mounted devices
Compact Operating Instructions, 06/2023, A5E35134047-AB
F
Power connection
G
Analyzer module 1 (AM1)
H
Analyzer module 2 (AM2)
I
Flame arrestors
J
Signal cable connection
Ex px safety equipment
B.1 Overview
109