Siemens SIPROCESS GA700 Compact Operating Instructions page 97

Continuous gas analytics
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Table 7-26
CALOMAT 7: Gas inlet conditions
Gas inlet conditions
Sample gas pressure
Pressure drop between sample gas inlet and out‐
let
Sample gas flow
Sample gas temperature and humidity
Table 7-27
CALOMAT 7: Temperature of sample chamber and gas path
Temperature of sample chamber and gas path
Gas path
All versions
Table 7-28
CALOMAT 7: Climatic conditions
Climatic conditions
Storage and transport
Max. permissible ambient temperature surround‐
ing the basic device during operation
Ambient humidity (relative humidity) during stor‐
age, transport or operation
Restriction for installing an ULTRAMAT 7 analyzer module (combined operation): 5 ... 45 °C (41 ... 113 °F)
1)
Avoid condensation on the installed components
2)
Table 7-29
CALOMAT 7: Gas connections
Gas connections
Connection fittings
Table 7-30
CALOMAT 7: Parts in contact with sample gases
Materials of wetted parts
Gas connection
Clamping rings and union nut (set)
Sample gas pipes
Sensor mounting block
Sensor
Gasket, contained in the sensor module
SIPROCESS GA700 rack and wall mounted devices
Compact Operating Instructions, 06/2023, A5E35134047-AB
-300 to +400 hPa
< 50 hPa at 1.5 l/min
30 to 90 l/h (0.5 to 1.5 l/min)
0 ... 70 °C (32 ... 158 °F) with relative humidity < 90%, avoid condensation
in the gas path (moisture content results in measurement errors)
5 K above ambient temperature
72 °C (162°F)
- 30 ... +70 °C (-22 ... 158 °F)
0 ... 50 °C (32 ... 122 °F)
 1)
< 90%
2)
Pipe diameter: Ø 6 mm (0.23") (outer diameter)
Stainless steel (Material No. 1.4571)
Stainless steel (Material No. 1.4401)
Stainless steel (Material No. 1.4404)
Stainless steel (Material No. 1.4571)
Si, SiO
Perfluorelastomere FFPM
N
, Au, epoxy resin, glass
x
y
Technical data
7.1 General technical specifications
97