Connecting The Signal Cables - Siemens SIPROCESS GA700 Compact Operating Instructions

Continuous gas analytics
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Connecting
4.7 Connecting the wall-mounted device
4.7.4.2

Connecting the signal cables

Overview
Figure 4-11
Procedure
1. Undo the six screws
2. Remove the shielding plate
3. Strip the end of signal cable to a length of approx. 1 cm.
4. Loosen the union nut from the respective cable gland
mounted device.
5. Insert the signal cable through this cable gland
62
Door screws
Shielding plate
Cable glands for signal lines (7 units)
Wall-mounted device, connecting signal cables
and open the door of the wall-mounted device.
on the right side panel. To do this loosen the screws.
.
Compact Operating Instructions, 06/2023, A5E35134047-AB
on the underside of the wall-
SIPROCESS GA700 rack and wall mounted devices