Install Nir Upgrade For Classic; Remove The Test Chamber - Juniper HarvestMaster GrainGage H2 NIR Upgrade Installation Manual

Table of Contents

Advertisement

H2 NIR Upgrade Installation Guide

2 Install NIR Upgrade for Classic

2.1 Remove the Test Chamber

1. Turn off air and power. Follow all lockout/tagout procedures.
notch
Install NIR Upgrade for Classic
2. Check the bottom back crossbar. If the crossbar has no
notch, replace with the notched version of 26242 H2M
Crossbar.
Note: The notch is positioned on the inside of the
GrainGage facing the bottom slide gate to allow more
clearance between the crossbar and an open bottom
slide gate.
3. At the T fittings, disconnect the air lines that connect
the air valve to the test chamber.
4. Shorten the air lines to 16 in.
Disconnect the two air
lines and shorten to 16 in.
notch
19

Advertisement

Table of Contents
loading