H2 NIR Upgrade Installation Guide
Install NIR Upgrade for Classic
lift the chamber to check for interference with the fill
guide. With the chamber lifted, look up through the
test chamber and center it around the fill guide. Then,
tighten the test chamber screws completely.
9. If you adjusted the load cell limit screw (in step 5), use
a 5/16 wrench and adjust it to have a 0.006 in. gap.
Load cell overload
protection screw
10. Attach the limit switch sensor to the air cylinder
leaving a 1/8–1/4 in. gap between the end fitting and
the band.
Correct position of the
limit switch sensor on
the air cylinder
25