H2 NIR Upgrade Installation Guide

1.7 Connect to Mirus

1. Download and install Mirus 4.5.0 or above.
Note: When asked to launch Mirus, do NOT launch.
2. Enable the H3 Plugin in the Plugin Manager.
3. Install the SCiO Service for the sensor.
Note: Each sensor has its own activation link. If you have not yet received your link,
a. Go to harvestmaster.com/support. Tap My Product.
b. Tap SCiO License Renewal.
4. Open Mirus and connect the H3 GrainGage plugin.
If Mirus does not successfully connect to the H3 GrainGage, complete the following steps in order.
a. Press the red button on the system controller to power off the GrainGage.
b. Close Mirus.
c. Unplug the CAN to USB cable from the computer.
d. Unplug the serial to USB cable from the computer.
e. Check that the SCiO service is running by checking the Windows Services application, as shown
below.
f. Plug the serial to USB cable into the computer.
g. Press the red button on the system controller to turn on the GrainGage.
h. Plug the serial to USB cable into the computer.
i. Open Mirus.
j. Tap the H3 icon to connect to the GrainGage.
Install NIR Upgrade for Single, Twin, and Triple