5 - 2
mAs Relay Adjustment
•
Select mAs Relay Adjust in the "Polydoros .. Service\Adjustment\" window.
•
Connect the mAs meter to D220.X16 (mAs test socket).
•
Exit the message window with OK..
NOTICE
•
Select "Exposure ON".
•
Enter the value measured on the mAS-meter in "Measured value: ..." using the arrow
keys.
•
Exit the window with OK.
•
Disconnect the mAs meter and reinstall the mAs jumper.
Default for SDM
(For the instant stored image option only, Item No. 47 82 517 J1042)
•
Select the "Polydoros .. -Service/Adjustment/Iontomat Sensitivity" window.
Program + 1.3 (default for SDM)
•
Exit the window with OK.
NOTICE
LITHOSTAR Multiline
System Manual
Perform the mAs relay adjustment:
1. During startup
2. After replacing the D100
3. If there are deviations in mAs values in "mAs Mode".
The data may be updated in the XCU with "Transfer Value"
allowing test exposures to be made without exiting the service
software window.
RXL2-120.034.05
Rev. 02
Page 2 of 2
12.99
TD PS 24
Basic Settings
Siemens AG
Medical Engineering
5
5