2 | DX-D 40C, DX-D 40G | Contents Contents Legal Notice ................4 Introduction to this Manual ........... 5 Scope .................6 Documentation by the manufacturer ....6 Warnings, Cautions, Instructions and Notes ....7 Disclaimer ..............8 Introduction to the DR Detector ..........9 Intended Use ............
Page 3
DX-D 40C, DX-D 40G | Contents | 3 Regular Inspection and Maintenance ... 43 Replacement Parts Support ......44 Repair ............45 Environmental Protection ........46 Safety Directions ............48 DR Detector Battery ........51 Safety directions for the power supply ..53...
Agfa and the Agfa rhombus are trademarks of Agfa-Gevaert N.V., Belgium or its affiliates. DX-D 40G and DR 40C are trademarks of Agfa HealthCare N.V., Belgium or one of its affiliates. All other trademarks are held by their respective owners and are used in an editorial fashion with no intention of infringement.
DX-D 40C, DX-D 40G | Introduction to this Manual | 5 Introduction to this Manual Topics: • Scope • Warnings, Cautions, Instructions and Notes • Disclaimer 0290A EN 20140626 1622...
6 | DX-D 40C, DX-D 40G | Introduction to this Manual Scope This manual contains information for the safe and effective operation of the DX-D 40C and DX-D 40G wireless DR Detectors and peripheral equipment, further referred to as the DR Detector.
DX-D 40C, DX-D 40G | Introduction to this Manual | 7 Warnings, Cautions, Instructions and Notes The following samples show how warnings, cautions, instructions and notes appear in this document. The text explains their intended use. Warning: Warnings are directions which, if they are not...
8 | DX-D 40C, DX-D 40G | Introduction to this Manual Disclaimer Agfa assumes no liability for use of this document if any unauthorized changes to the content or format have been made. Every care has been taken to ensure the accuracy of the information in this document.
DX-D 40C, DX-D 40G | Introduction to the DR Detector | 9 Introduction to the DR Detector Topics: • Intended Use • Intended User • Configuration • Equipment Classification • Options and Accessories • Operation Controls • System Documentation •...
10 | DX-D 40C, DX-D 40G | Introduction to the DR Detector Intended Use The DR Detector is a wired and wireless radiographic digital X-ray imaging device commonly referred to as flat panel detector. It is designed for all general radiography applications. The DR Detector will be used in a radiological environment by qualified staff to capture the X-ray images and send these images to the imaging processing software.
DX-D 40C, DX-D 40G | Introduction to the DR Detector | 11 Configuration The DR Detector is a component that can be integrated in an X-ray system, connected to a workstation and to the X-ray generator. DR Detector DR Detector battery...
12 | DX-D 40C, DX-D 40G | Introduction to the DR Detector Equipment Classification Per EN/IEC60601-1, Medical Electrical Equipment, General Requirements for Safety 3rd Edition, the DR Detector, including the battery pack, is classified as following. Class I Internally powered...
DX-D 40C, DX-D 40G | Introduction to the DR Detector | 13 Options and Accessories • DR Detector battery • DR Detector battery charger • Handle with screws • Click-on grid Recommended grid specifications Size 14 inch x 17 inch...
14 | DX-D 40C, DX-D 40G | Introduction to the DR Detector Operation Controls Topics: • DX-D 40C, DX-D 40G • DR Detector Battery Charger • System Control Unit • DR Detector Switch on the NX Workstation 0290A EN 20140626 1622...
DX-D 40C, DX-D 40G | Introduction to the DR Detector | 15 DX-D 40C, DX-D 40G DR Detector battery lock lever Unlock the battery. DR Detector battery Supplies power to the detector during wireless communication. Antenna of the wireless network adapter Operation in wireless configuration.
16 | DX-D 40C, DX-D 40G | Introduction to the DR Detector DR Detector Battery Charger The battery charger has three slots to insert a battery. Figure 3: DR Detector Battery Charger Warning: Do not use the battery charger within the patient’s vicinity.
DX-D 40C, DX-D 40G | Introduction to the DR Detector | 17 System Control Unit The System Control Unit is connected to the DR Detector via wireless network. Power switch Status indicator • Blinking green: starting up • Green: ready •...
18 | DX-D 40C, DX-D 40G | Introduction to the DR Detector DR Detector Switch on the NX Workstation The DR Detector Switch is available in the title bar of the NX application. The DR Detector Switch shows which DR Detector is active and shows its status.
DX-D 40C, DX-D 40G | Introduction to the DR Detector | 19 System Documentation The documentation consists of a User manual (this document) and related documentation: • NX User Manual (4420). • NX Key User Manual (4421). • NX Getting Started Sheets (4424).
20 | DX-D 40C, DX-D 40G | Introduction to the DR Detector Training The user must have received adequate training on the safe and effective use of the system before attempting to work with it. Training requirements may vary from country to country. The user must make sure that training is received in accordance with local laws or regulations that have the force of law.
DX-D 40C, DX-D 40G | Introduction to the DR Detector | 21 Product Complaints Any health care professional (for example a customer or a user) who has any complaints or has experienced any dissatisfaction with the quality, durability, reliability, safety, effectiveness, or performance of this product must notify Agfa.
22 | DX-D 40C, DX-D 40G | Introduction to the DR Detector Compatibility The system must only be used in combination with other equipment or components if these are expressly recognized by Agfa as compatible. A list of such equipment and components is available from Agfa service on request.
DX-D 40C, DX-D 40G | Introduction to the DR Detector | 23 Compliance Documentation by the manufacturer For complete safety directions and technical documentation, refer to the ViVIX-S 1417W User Manual that is delivered with the product. 0290A EN 20140626 1622...
24 | DX-D 40C, DX-D 40G | Introduction to the DR Detector Connectivity Topics: • Wireless Communication • Wired communication 0290A EN 20140626 1622...
DX-D 40C, DX-D 40G | Introduction to the DR Detector | 25 Wireless Communication Wireless communication is established between the internal wireless module of the DR Detector and the NX workstation via the wireless access point. The DR Detector is compliant with IEEE 802.11n (2.4 GHz/5 GHz). The available frequency band varies depending on local radio laws and system requirements.
26 | DX-D 40C, DX-D 40G | Introduction to the DR Detector Wired communication The use of accessories and cables other than those specified or sold by the manufacturer as replacement parts, may result in increased radiation emissions or decreased stability of the equipment.
DX-D 40C, DX-D 40G | Introduction to the DR Detector | 27 Installation Installation and configuration is performed by an Agfa trained and authorized service engineer. Contact your local support organization for more information. On a configuration with multiple DR Detectors of the same type, it is required to apply labeling to the DR Detector containing a unique nickname for each DR Detector.
Page 28
28 | DX-D 40C, DX-D 40G | Introduction to the DR Detector Note: This product may malfunction due to electromagnetic waves caused by portable personal telephones, transceivers, radio- controlled toys, etc. Be sure to avoid having objects such as these, which affect this product, brought near the product.
DX-D 40C, DX-D 40G | Introduction to the DR Detector | 29 Messages Under certain conditions the DR Detector shows a dialog box containing a message in the middle of the screen of the NX workstation. This message informs the user that either a problem has occurred or that a requested action cannot be performed.
30 | DX-D 40C, DX-D 40G | Introduction to the DR Detector Labels Symbol Explanation The system identification can be accessed by removing the battery. Topics: • DR Detector identification label • Documentation by the manufacturer 0290A EN 20140626 1622...
DX-D 40C, DX-D 40G | Introduction to the DR Detector | 31 DR Detector identification label Label Meaning Writable label to identify and dedicate a DR Detector to an X- ray system bucky. 0290A EN 20140626 1622...
32 | DX-D 40C, DX-D 40G | Introduction to the DR Detector Documentation by the manufacturer For complete safety directions and technical documentation, refer to the ViVIX-S 1417W User Manual that is delivered with the product. 0290A EN 20140626 1622...
DX-D 40C, DX-D 40G | Introduction to the DR Detector | 33 Cleaning and Disinfecting All appropriate policies and procedures should be followed to avoid contamination of the staff, patients and equipment. All existing universal precautions should be extended to avoid potential contaminations and to avoid patients coming into (close) contact with the device.
34 | DX-D 40C, DX-D 40G | Introduction to the DR Detector Cleaning To clean the exterior of the equipment: 1. Stop the system Warning: When the equipment is going to be cleaned, be sure to turn OFF the power of each device, and to unplug the power cord from the AC outlet.
DX-D 40C, DX-D 40G | Introduction to the DR Detector | 35 Use of protective plastic bag Warning: Liquids ingressing the DR Detector may cause malfunction and contamination. If there is a chance that the detector comes in contact with liquids (bodily fluids, disinfectants,...), the DR Detector must be wrapped in a protective...
36 | DX-D 40C, DX-D 40G | Introduction to the DR Detector Disinfecting To disinfect the device, use only disinfectants and disinfection methods that are approved by Agfa and that correspond to the national regulation and guidelines. If you plan to use other disinfectants, approval of Agfa is needed before use, as most disinfectants can damage the device.
DX-D 40C, DX-D 40G | Introduction to the DR Detector | 37 Approved disinfectants Refer to the Agfa website for specifications on the disinfectants that have been found compatible with the cover material of the device and can be used on the outer surface of the device.
38 | DX-D 40C, DX-D 40G | Introduction to the DR Detector Safety directions for disinfection Warning: When the equipment is going to be cleaned, be sure to turn OFF the power of each device, and to unplug the power cord from the AC outlet. Never use anhydrous or high solvency alcohols, benzine, thinner or any other flammable cleaning agent.
Page 39
DX-D 40C, DX-D 40G | Introduction to the DR Detector | 39 Warning: Do not reuse wipes. Caution: Disinfecting solution or wipes may cause eye and skin irritation. Wear gloves and wash hands with soap and water following use. Consult the manufacturer’s Material Safety Data Sheets (MSDS) and recommendations on the product label for additional information prior to use.
40 | DX-D 40C, DX-D 40G | Introduction to the DR Detector Maintenance Always consult the Agfa Service documentation and an Agfa trained and authorized service engineer for complete maintenance schedules. In order to ensure that the equipment is used safely and normally, be sure to inspect the equipment before use.
DX-D 40C, DX-D 40G | Introduction to the DR Detector | 41 Daily inspection Warning: For safety reasons, be sure to turn OFF the power to each piece of equipment before performing the following. Otherwise, an electric shock may result.
42 | DX-D 40C, DX-D 40G | Introduction to the DR Detector Half-yearly inspection To indicate when the half-yearly calibration is due, a message is displayed on the NX workstation. Perform calibration half-yearly or when exposure conditions have changed significantly. For details, refer to the DX-D DR Detector Calibration Key User Manual (0134).
DX-D 40C, DX-D 40G | Introduction to the DR Detector | 43 Regular Inspection and Maintenance In order to ensure the safety of patients, operating personnel and third parties, and to maintain the performance and reliability of the equipment, be sure to perform regular inspection at least once a year.
44 | DX-D 40C, DX-D 40G | Introduction to the DR Detector Replacement Parts Support Parts required to maintain the functioning of the product will be stocked for seven years after discontinuance of production, to allow for repair. 0290A EN 20140626 1622...
46 | DX-D 40C, DX-D 40G | Introduction to the DR Detector Environmental Protection Disposal of this product in an unlawful manner may have a negative impact on health and on the environment. Therefore, when disposing of this product, be absolutely sure to follow the procedure which is in conformity with the laws and regulations applicable in your area.
Page 47
DX-D 40C, DX-D 40G | Introduction to the DR Detector | 47 For battery replacements please contact your local sales organization. 0290A EN 20140626 1622...
48 | DX-D 40C, DX-D 40G | Introduction to the DR Detector Safety Directions Warning: Strictly observe all warnings, cautions, notes and safety markings within this document and on the product. Warning: Safety is only guaranteed when an Agfa certified field service engineer has installed the product.
Page 49
DX-D 40C, DX-D 40G | Introduction to the DR Detector | 49 equipment, or if liquid is spilled, it may result in fire or electric shock. Warning: Do not hit or drop the equipment. The equipment may be damaged if it receives a strong jolt, which may result in fire or electric shock if the equipment is used without being repaired.
Page 50
50 | DX-D 40C, DX-D 40G | Introduction to the DR Detector Caution: Be sure to use the detector on a flat surface so it will not bend. Otherwise, the internal image sensor may be damaged. Be sure to securely hold the detector while using it in upright positions.
DX-D 40C, DX-D 40G | Introduction to the DR Detector | 51 DR Detector Battery Safety Directions Warning: Do not use any charger other than that specifically provided for use with the equipment. Use only a power adapter complying with IEC 60601-1 or IEC 60950-1.
Page 52
52 | DX-D 40C, DX-D 40G | Introduction to the DR Detector Avoid dropping or subjecting the product to severe impacts. To avoid the risk of injury, do not touch the internal parts of the battery if it has been cracked or otherwise damaged.
DX-D 40C, DX-D 40G | Introduction to the DR Detector | 53 Safety directions for the power supply Warning: Do not operate the equipment using any type of power supply other than the one indicated on the rating label. Otherwise, it may result in fire or electric shock.
Page 54
54 | DX-D 40C, DX-D 40G | Introduction to the DR Detector Caution: When using the power supply, care must be taken to ensure that there is either a mains plug or an all-cable disconnecting device in the internal installation fitted near the device and that it is easily accessible in case of emergency.
DX-D 40C, DX-D 40G | Introduction to the DR Detector | 55 Safety directions for the System Control Unit Warning: Do not block the ventilation ports to prevent overheating. Overheating can cause system malfunction and damages. Warning: Ensure continuous power supply to the system, with voltage and current according to the product specifications.
56 | DX-D 40C, DX-D 40G | Getting started Getting started Topics: • Starting the DR Detector • Basic Workflow DR Detector • Guidelines for Pediatric Applications • Stopping the DR Detector 0290A EN 20140626 1622...
DX-D 40C, DX-D 40G | Getting started | 57 Starting the DR Detector Caution: Be sure to use only the dedicated battery pack for the DR Detector. To start the DR Detector: 1. Fully charge the battery. Charge the battery on the day of examination or on the previous day.
Page 58
58 | DX-D 40C, DX-D 40G | Getting started Note: Before operating the detector, start up the NX workstation. Press and hold the power button for 1 second. Figure 7: Power button After startup the green status indicator is lit, indicating the power status.
60 | DX-D 40C, DX-D 40G | Getting started Step 1: retrieve the patient info At the NX workstation: 1. When a new patient comes in, define the patient info for the exam. 2. Start the exam. Step 2: select the exposure 1.
DX-D 40C, DX-D 40G | Getting started | 61 Step 3: prepare the exposure In the examination room: 1. Position the DR Detector. When using the bucky, check that the identification labels on the DR Detector and on the bucky match. Do not use a DR Detector that is dedicated to another bucky.
62 | DX-D 40C, DX-D 40G | Getting started Step 4: check the exposure settings On the DR Detector Switch: 1. Check if the DR Detector Switch displays the name of the DR Detector that's being used 2. If a wrong DR Detector is displayed, select the right DR Detector by clicking the drop down arrow on the DR Detector Switch.
DX-D 40C, DX-D 40G | Getting started | 63 Step 5: execute the exposure Press the exposure button to execute the exposure. Instruction: Make sure the generator is ready for exposure before you press the exposure button. Warning: The radiation indicator on the control console lights up during exposure release.
64 | DX-D 40C, DX-D 40G | Getting started Positioning the DR Detector When performing an exposure, keep in mind the following detector orientation aids: • tube side • patient orientation marker Figure 8: Detector orientation aids 1. Tube side of the detector 2.
Page 65
DX-D 40C, DX-D 40G | Getting started | 65 2. Patient orientation (AP) 3. Result on monitor Table 2: Chest PA Landscape Figure 10: Chest PA landscape 1. Detector orientation (Landscape) 2. Patient orientation (PA) 3. Result on monitor Table 3: Table with bucky...
Page 66
66 | DX-D 40C, DX-D 40G | Getting started Table 4: Wallstand bucky Wallstand with left loading bucky, portrait Wallstand with left loading bucky, landscape Wallstand with right loading bucky, portrait Wallstand with right loading bucky, landscape 0290A EN 20140626 1622...
DX-D 40C, DX-D 40G | Getting started | 67 Guidelines for Pediatric Applications Caution: Children are more radiosensitive than adults. Adopting the Image Gently campaign guidelines and reducing dose for radiographic procedures while maintaining acceptable clinical image quality will benefit patients.
Page 68
68 | DX-D 40C, DX-D 40G | Getting started • Avoid multiple scans and use alternative diagnostic studies (such as ultrasound or MRI) when possible. 0290A EN 20140626 1622...
DX-D 40C, DX-D 40G | Getting started | 69 Stopping the DR Detector To stop the DR Detector: 1. Turn off the detector. Press and hold the power button (approx. 3 seconds). All the status indicator lights are off. 2. While holding down the battery pack, slide the lock levers toward (unlock) (1), put your fingers on the battery edge that lifts up, and then pull the edge to remove the battery pack (2).
DX-D 40C, DX-D 40G | Advanced Operating | 71 Detector Status Indicators Table 5: Detector status Status Green status Orange status Blue status indicator indicator indicator (Power) (Active) (Data) Power on Detector ready Detector sleep mode Data communication ongoing Wireless data...
72 | DX-D 40C, DX-D 40G | Advanced Operating Charging a battery To charge a battery using the battery charger: 1. Connect the power supply to the mains power and to the power socket of the battery charger. 2. Insert the battery in an empty slot of the battery charger.
74 | DX-D 40C, DX-D 40G | Problem solving Artifact in DR Detector images Details An artifact is visible in the images produced by a DR Detector. Cause Exposure conditions have changed significantly since latest calibration. Brief Solution Perform calibration of the DR Detector.
DX-D 40C, DX-D 40G | Problem solving | 75 DX-D 40C, DX-D 40G Symptom Cause Remedy The detector will The battery is not Attach the batter. not turn on. attached. The battery pack is not Fully charge the battery pack.
Page 76
76 | DX-D 40C, DX-D 40G | Problem solving Symptom Cause Remedy consumed For purchase of consumables, quickly. contact your sales representative or local dealer. The battery was In low temperatures the battery charged or used in low capacity decreases. Use a temperatures.
Page 79
DX-D 40C, DX-D 40G | Technical Data | 79 Altitude max. 2000 m Warming-up time 30 minutes Dimensions Dimensions approx. 384 x 460 x 15.5 mm width x length x height < 3.4 kg Weight (incl. battery) Maximum load 100 kg on an area of 40 mm in...
82 | DX-D 40C, DX-D 40G | Technical Data System Control Unit System Control Unit Model number FXRS-03A Manufacturer Vieworks Co., Ltd. #107-108, 601-610, Suntechcity 2, 52 Sagimakgol-ro (307-2, Sangdaewon- dong), Jungwon-gu, Seongnam-si, Gyeonggi-do, 462-736 South Korea Rated power supply AC100 to 240V, 50/60 ㎐, Max.
DX-D 40C, DX-D 40G | Remarks for HF-emission and immunity | 83 Remarks for HF-emission and immunity Documentation by the manufacturer For complete safety directions and technical documentation, refer to the ViVIX-S 1417W User Manual that is delivered with the product.