DX-D 40C, DX-D 40G

Symptom
The detector will
not turn on.
The status
indicator of the
System Control
Unit does not
light up.
The status
indicator of the
System Control
Unit does not
light up in
green.
The green status
indicator lights
up and the
orange and blue
status indicators
are flashing.
All status
indicators are
flashing.
Two status
indicators are
flashing and the
third is flashing
slowly.
A fully charged
battery is
Cause
The battery is not
attached.
The battery pack is not
charged.
The power cord is
unplugged from the AC
outlet.
A hardware error has
occurred.
An error has occurred
during registration of
the DR Detector.
An error has occurred
during data
communication.
A hardware error has
occurred.
The battery capacity
decreases.
DX-D 40C, DX-D 40G | Problem solving | 75
Remedy
Attach the batter.
Fully charge the battery pack.
Connect the plug to the AC
outlet firmly. If it still does not
work, replace the System
Control Unit.
Turn off the System Control
Unit and turn it on again. If it
still does not work, replace the
System Control Unit.
Check the network connection
on the System Control Unit.
Check the network
configuration of the
workstation.
Check if the System Control
Unit is turned on.
Check if the wireless network
communication is stable.
Turn off the DR Detector and
turn it on again.
The DR Detector battery can
deteriorate because of its
characteristics and structure.
0290A EN 20140626 1622