62 | DX-D 40C, DX-D 40G | Getting started

Step 4: check the exposure settings

On the DR Detector Switch:
1. Check if the DR Detector Switch displays the name of the DR Detector
that's being used
2. If a wrong DR Detector is displayed, select the right DR Detector by
clicking the drop down arrow on the DR Detector Switch.
3. Check the DR Detector Status icon.
On the X-ray system:
1. Check if the exposure settings displayed on the console are suitable for the
exposure.
2. Check if no error messages are displayed on the X-ray system.
Automatic exposure detection
The DR detector detects X-ray exposure to automatically perform the image
acquisition.
Before performing the exposure, the DR detector must be ready. Check the
status of the DR detector in the DR Detector Switch.
0290A EN 20140626 1622
Warning: The center of the DR Detector must be in the
exposed area. Positioning the center of the DR Detector
outside the exposed area can cause failure to trigger the
image acquisition.
Warning: Do not hit or drop the equipment. If it receives a
strong jolt, image acquisition can be triggered without X-ray
exposure.
Warning: Specific exposure conditions (use of grid,
thickness of the exposed object) can cause failure to trigger
the image acquisition or horizontal artifacts in the acquired
image.
Caution: Excessive ambient temperature may impact
performance of DR Detectors and cause permanent damage to
the equipment. If ambient temperature and humidity is outside
the range of 10 - 35 °C and 30 - 85% RH, do not operate the
system or use air conditioning. Warranty will be void if it is
obvious that operating conditions are not met.