Step 3: prepare the exposure

In the examination room:
1. Position the DR Detector.
When using the bucky, check that the identification labels on the DR
Detector and on the bucky match. Do not use a DR Detector that is
dedicated to another bucky.
2. Position the patient.
Apply radiation protective measures for the patient if needed.
3. Check if the X-Ray system position is suitable for the exposure.
4. Position the X-Ray tube with respect to the DR Detector and the patient.
5. Set the correct distance between DR Detector and X-Ray tube.
6. Switch on the light on the collimator. Adapt collimation if required.
Take care that the collimated area is not larger than the detector.
Warning: Monitor the patient position (hands, feet, fingers,
etc.) with special care to avoid injury to the patient caused
by unit movements. Patient hands must be kept away from
mobile components of the unit. Intravenous tubing,
catheters and other patient connected lines should be
routed away from moving equipment.